HHS 100 Series Moisture Analyzer

---

 

 

 

High Temperatures Capability

 

The HHS 100 series sensor is designed for direct insertion into exhaust ducts and stacks from boilers, driers, or ovens. A constant sample of the process stream is measured to ensure accuracy.  The sensor determines % volume water and dry-bulb temperature of samples.  The HHS 100 series has an advantage compared to other sensors because it can operate under much higher temperatures ranging from 0 to 400 degree Celsius.

 

NiMH Rechargeable Supply

 

Most stack measurements must be made in difficult locations- on the roof of a building, or at a high level about an oven.  Mains-powered instruments are always a major disadvantage in these conditions.  Significant time can be wasted arranging for suitable extension leads and/or weatherproof power supplies.  High capacity rechargeable NiMH batteries ensure that the HHS 100 has an effective measurement time of at least 3 hours on one set of four standard AA cells.  No external supply or extension leads are necessary.  Additional sets of rechargeable batteries can be supplied for extended measurement times.

 

LCD Display

 

An easy to read LCD screen displays the measured values of % volume water and dry-bulb temperature.  Display of dry-bulb temperature is user-selectable, in either Celsius or Fahrenheit.

 

 

 

Click on the picture to see a larger view

 

 

 

                          Overall Performance Specifications

 

Temperature Range

 

   Sample Probe

32 to 752 degree F (0 – 400 degree C)

   Electronic enclosure

32 to 104 degree F (0 to 40 degree C)

Moisture Range

 

   Sample temp(greater than 85C)

0 to 80% vol. water

   Sample temp(less than 85C)

0 to 100% vol. water

Response time, 1/e

6 seconds

Warm-up time

5 minutes, typical

Battery Life

3 hours, typical

 

 

Part Number

Accessory

HHS100

Moisture analyzer kit

HHS100-01

Hand-held analyzer

HHS100-02

Instruction Manual

HHS100-03

Protective carrying case

HHS100-04

Set of NiMH batteries

HHS100-05

NIST calibration certificate

 

 

 

 

 

 

 

 

 

 

 

 

 

 

---

 

 Site Links

 

 

 

 

---

 

AMT, Inc.

560 W. 3rd St. • Jamestown, NY 14701 • (716) 488-9827 • Fax: (716) 664-2898

 

Copyright © AMT, Inc. 2005 – 2006